1 Select the setup key
2 Select Alarm > FHR.
3 Select a value from 60 ~ 205 for Low Alarm Limit.
4 Select a value from 65 ~ 210 for High Alarm Limit.
5 Select OK.

7.2.5 Changing FHR Alarm Delay

You can change the FHR alarm delay. The alarm delay indicates how long the measured result
continues exceeding its limit before the alarm is triggered.
1 Select the setup key
2 Select Alarm > FHR1 > Alarm Delay.
3 Select a value from 0 ~ 300.
4 Select OK.
7.3 Monitoring FHR with DECG
NOTE:
This feature is only available on F6.

7.3.1 Contraindications

The fetal spiral electrode can be used when amniotic membranes are adequately ruptured and
sufficient cervical dilatation is ensured. The fetal electrode tip is designed to penetrate the
epidermis of the fetus; therefore, trauma, hemorrhage and/or infection can occur. The electrode
should be used with strict adherence to aseptic technique.
The fetal spiral electrode should not be applied to the fetal face, fontanels or genitalia.
Do not apply the fetal spiral electrode when placenta previa is present; when the mother has
visible genital herpes lesions or reports symptoms of prodromal lesions; when the mother is HIV
sero-positive; when mother is a confirmed carrier of hemophilia and the fetus is affected or of
unknown status; or when it is not possible to identify fetal presenting part where application is
being considered. This method is not recommended when fetus is extremely premature, or in the
presence of a maternal infection such as Hepatitis B, Group B hemolytic strep, syphilis or
gonorrhea, unless a clear benefit to the fetus or mother can be established.

7.3.2 Parts Required

1) DECG cable
2) Fetal spiral electrode
on the main interface.
on the main interface.
(F6)
3) Disposable attachment pad electrode