EDAN F6 User Manual page 32

Fetal & maternal monitor
Hide thumbs Also See for F6:
Table of Contents

Advertisement

F6, F6 Express Fetal & Maternal Monitor User Manual
b)
: Alarm reviewing key. Select this key to open the alarm reviewing menu.
c)
: Paper advancing key. Select this key to advance the paper for 8 cm (PHILIPS
paper) or 7 cm (GE paper).
d)
: Display mode switch. F6 Express monitor has three display modes: maternal-fetal
display mode, fetal display mode and maternal display mode. Select this key, and the display
mode will switch to the next one in order.
e)
: Mat. Info key. Select this key to open maternal information menu for inputting or
changing the patient's ID and name.
f)
: Setup key. Select this key to open setup main menu.
(2) Trace/Menu Window
The trace/menu window occupies most space of the screen. During monitoring or reviewing, it
displays traces; during setting, it displays setup menus.
The background pane bar supports two standards: 30 ~ 240 (American standard) and 50 ~ 210
(International standard).
The green band in between the fetal heart rate panes indicates the preset alarm range (the top
edge is not higher than 180 and the bottom edge is not lower than 100). It makes it easy to
observe if the FHR exceeds the normal range. So you can easily tell if the fetal heart rate is too
low or too high.
(3) Numeric Window
The fetal monitoring numerics are displayed here.
(4) Status Window
g)
Power indicator
- AC power supplied.
- no AC power supplied.
h)
Battery indicator
- battery is loaded; the green pane indicates the charge of the battery.