EDAN F6 User Manual page 79

Fetal & maternal monitor
Hide thumbs Also See for F6:
Table of Contents

Advertisement

F6 and F6 Express Fetal & Maternal Monitor User Manual
DECG LEADS OFF
DECG UNPLUGGED
DECG SIGNAL LOSS
DECG EQUIP MALF
Signals
Overlap
(FHR1, FHR2)
Signals
Overlap
(DFHR, FHR2)
The spiral electrode is not well
DECG
connected.
The DECG lead is not well
DECG
connected to the monitor.
DECG signal is too weak for
DECG
the system to analyze.
The DECG board can not
DECG
communicate with the system
successfully.
US transducer 1 and US
transducer 2 are aimed at the
US
same fetal heart; the signals
overlap.
US transducer 1 is aimed at
US +
the
fetus
electrode is attached to; the
DECG
signals overlap.
that
the
spiral
- 71 -
Fetal Monitoring Display
Check the connection of the
spiral electrode.
Check the connection of the
DECG cable.
Check if the spiral electrode
is well attached to the fetus;
check
the
patient's
condition.
Restart the monitor and try
again,
contact
manufacturer
if
connection still fails.
Adjust
one
of
the
transducers until another
fetal
heart
signal
detected.
Adjust the US transducer
until another fetal heart
signal is detected.
the
the
US
is