Detector - Agilent Technologies 6490 Manual Book

6400 series triple quadrupole lc/ms system
Hide thumbs Also See for 6490:
Table of Contents

Advertisement

3
Agilent Triple Quadrupole MS and Sensitivity

Detector

Detector

Figure 30
Detector components
62
The High Energy Dynode detector assembly is unique to Agilent
(Figure
30).
The two dynodes are orthogonal to the ion beam and neutrals.
This orientation reduces the possibility of neutral molecules
impacting the detector while at the same time attracting the
ions with high voltages. The dynodes serve to convert the ions
to electrons before they impact the multiplier. The off-axis
design also allows neutrals to pass through without hitting the
detector.
The multiplier has a long lifetime since only electrons are
allowed to impact it. Ions never impact its surface.
Agilent 6400 Series Triple Quad LC/MS Concepts Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

643064206460

Table of Contents