Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS
2

Innovative Enhancements in the 6490 Triple Quadrupole

Circled areas indicate enhancements
Figure 8
Innovative Enhancements in the 6490
The quadrupole consists of four parallel hyperbolic rods
through which selected ions are filtered before reaching
a collision cell where they are fragmented. The collision cell is
typically called the second quadrupole, but in the case of the
6460, geometrically it is actually a hexapole filled with nitrogen,
the same gas used in the ion source. In the 6490, the collision
cell is a hexapole field axial focusing curved collision cell.
The fragment ions formed in the collision cell are then sent to
the third quadrupole for a second filtering stage to enable
a user to isolate and examine multiple precursor to product ion
transitions (MRMs).

Innovative Enhancements in the 6490 Triple Quadrupole

The iFunnel Technology encompasses the first three
enhancements to the 6490: a hexabore capillary, the Dual Ion
Funnel technology and the Curved Collision Cell. The fourth
enhancement is improved quadrupole drive electronics.
Agilent 6400 Series Triple Quad LC/MS Concepts Guide