Inner Workings – Triple Quadrupole MS versus Single Quadrupole MS

Single quadrupole MS operation

Single quadrupole MS operation

Design for a single quadrupole mass spectrometer

m/z – mass/charge ratio
14
To better understand the specific hardware features of the
Agilent 6400 Series Triple Quadrupole Mass Spectrometer, this
section first reviews the fundamental aspects of the single
quadrupole mass spectrometer.
Mass spectrometry is based on the analysis of ions moving
through a vacuum.
The ionization of a sample occurs in the ion source that is
shown, schematically, on the left. The ions are analyzed by
a mass analyzer (mass filter) that controls the motion of the
ions as they travel to the detector to be converted into actual
signals.
Figure 1
Schematic for single quadrupole mass spectrometer
The quadrupole mass analyzer consists of four parallel rods to
which specific DC and RF voltages are applied. These rods filter
out all ions except those of one or more particular m/z values as
determined by the voltages applied.