Working with the Control Module

Troubleshooting

Configuring the LC System
The LC system is self-configuring to a large extent. It recognizes automatically
which modules are installed. The layout of the Analysis screen changes according to
the modules present. You can use the Configure button [F2] in the System screen to
configure various features such as the Auto-on for the system or Loading (a
method), and After Error Condition.
If an error event occurs, you can configure the LC system to load a specified method
or turn off the LC system using the After Error Condition setting.
Troubleshooting
Internal diagnostics continuously monitor the module's condition and record any
unusual events in an electronic logbook. For example, missing vials or leaking
solvent will signal errors and record the errors in the logbook together with the time
and date the errors occurred. The logbook is self-updating where the newest entry
replaces the oldest entry. If a printer is connected you can easily print out the
logbook.
Troubleshooting the Control Module
If your control module does not work correctly, disconnect the module CAN
connector from the rear of the Agilent 1100 Series module it is attached to and
reconnect it.
If the problem still remains, then
• power down all connected devices and computers and wait 1 minute and then
restart, or
• try to use just one Agilent 1100 Series module.
If the problem still remains, call Agilent Technologies.