4 Restart the control module using the Restart function via the m (menu) key (available
5 Load the method from the PC card. See section "Exchanging Methods With the PC
NOT E
Working with the Control Module

Time Programming

from the main screens).
Card" on page 41.
Time Programming
To time-program selected settings during the analysis you can create a timetable.
Using the Timetable screen, you can create a time based program that will
automatically control the pump, detector, column compartment and external
contacts.
In some cases the settings will change instantaneously from the initial value to the
value specified after a certain time in the timetable (e.g. wavelength). In other cases
(solvent composition) these changes take place dynamically, approaching the set
value in a step wise and linear manner.
The timetable becomes part of the current method when the method is saved.
The timetable is accessed by selecting the Timetable button [F2] in the Analysis
screen.
A timetable line can be inserted by pressing the Insert button [F7] and consists of the
following:
• Time
Set the time span between the instant of injection and the desired parameter
change.
• Module
Choose the module that controls the parameter you want to change.
• Setting
Select the parameter to be changed.
• Value
Enter the desired parameter value.
43