Table 9-4:
8
9
10
11
9.13

SMART Meter Verification measurements

The SMART Meter Verification test will make measurements of the coil resistance, coil
signature, and electrode resistance and compare these values to the values taken during
the sensor signature process to determine the sensor calibration deviation, the coil circuit
health, and the electrode circuit health. In addition, the measurements taken by this test
can provide additional information when troubleshooting the meter.
Coil circuit resistance
LOI menu path
The coil circuit resistance is a measurement of the coil circuit health. This value is
compared to the coil circuit resistance baseline measurement taken during the sensor
signature process to determine coil circuit health. This value can be continuously
monitored using continuous SMART Meter Verification.
Coil signature
LOI menu path
The coil signature is a measurement of the magnetic field strength. This value is compared
to the coil signature baseline measurement taken during the sensor signature process to
determine sensor calibration deviation. This value can be continuously monitored using
continuous SMART Meter Verification.
Reference manual
Continuous SMART Meter Verification Test Parameters
Parameter
Electrode Resistance
mA Expected
mA Actual
mA Deviation
Manual: Diagnostics > Advanced Diag > Meter Verif > Measure-
ments > Manual Measure > Coil Resist
Continuous: Diagnostics > Advanced Diag > Meter Verif > Meas-
urements > Continual Meas > Coil Resist
Manual: Diagnostics > Advanced Diag > Meter Verif > Measure-
ments > Manual Measure > Coil Inductnce
Continuous: Diagnostics > Advanced Diag > Meter Verif > Meas-
urements > Continual Meas > Coil Inductnce
Advanced Diagnostics Configuration
(continued)
LOI menu path ( Diagnostics > Varia-
bles > MV Results > Continual Res )
Electrode Res
4–20 mA Expect
4–20 mA Actual
AO FB Dev