4.4.3
Sensor to transmitter wiring
Integral mount transmitters
Integral mount transmitters ordered with a sensor will be shipped assembled and wired at
the factory using an interconnecting cable. Use only the factory supplied cable provided
with the instrument. For replacement transmitters use the existing interconnecting cable
from the original assembly. Replacement cables, if applicable, are available (see
Figure
Figure 4-4:
A.
B.
Remote mount transmitters
Cables kits are available as individual component cables or as a combination coil/electrode
cable. Remote cables can be ordered directly using the kit numbers shown in
Table
alternative. To order cable, specify length as quantity desired. Equal length of component
cables is required.
Examples:
Table 4-3:
Cable kit #
08732-0065-0001
(feet)
08732-0065-0002
(meters)
08732-0065-0003
(feet)
Reference manual
4-4).
Replacement interconnecting cables
A
Socket module 08732-CSKT-0001
IMS cable 08732-CSKT-0004
4-4, and
Table
4-5. Equivalent Alpha cable part numbers are also provided as an
25 feet = Qty (25) 08732-0065-0001
25 meters = Qty (25) 08732-0065-0002
Component cable kits - standard temperature (-20°C to 75°C)
Description
Kit, component cables,
Std temp (includes Coil
and Electrode)
Kit, component cables,
Std temp (includes Coil
and Electrode)
Kit, component cables,
Std temp (includes Coil
and I.S. Electrode)
Remote Transmitter Installation
B
Individual cable
Coil
Electrode
Coil
Electrode
Coil
Instrinsically Safe Blue
Electrode
Table
4-3,
Alpha p/n
2442C
2413C
2442C
2413C
2442C
Not available
31