PP Preselector Peak

PP Preselector Peak
Syntax
Figure 7-209
PP Syntax
Description
The PP command peaks the preselector in the Agilent 8561E/EC,
Agilent 8562E/EC, Agilent 8563E/EC, Agilent 8564E/EC and Agilent
8565E/EC or when using the spectrum analyzer with a preselected
external mixer. Make sure the entire frequency span is in high band
(greater than 2.9 GHz), set the desired trace to clear-write mode, place
a marker on a desired signal, then execute PP. The peaking routine
zooms to zero span, peaks the preselector tracking, then returns to the
original position. To read the new preselector peaking number, use the
PSDAC command. Commands following PP are not executed until after
the analyzer has finished peaking the preselector.
Preselector peaking is not available if the spectrum analyzer resolution
NOTE
bandwidth is less than or equal to 100 Hz.
Example
10
20
30
560
OUTPUT 718;"CF 3GHZ;SP 500KHZ;"
OUTPUT 718;"TS;MKPK HI;MKCF;TS;PP;"
END