SPAC CAL Sweep data problem finding "bucket 1" of the
ERR 356
span accuracy calibration sweep.
SPAC CAL Cannot find the "x" intersection for "bucket 1"
ERR 357
of the span accuracy calibration sweep.
SPAC CAL Cannot find "bucket 2" of the span accuracy
ERR 358
calibration sweep.
SPAC CAL Cannot find the "x" intersection for "bucket 2"
ERR 359
of the span accuracy calibration sweep.
SPAC CAL The start bucket correction is out of range.
ERR 360
SPAC CAL The percent of span correction is out of range.
ERR 361
Error codes 400 through 599 relate to IF failures. Instrument
service is required These errors are generated when the automatic
IF adjustment routine detects a fault. These errors are designed to
assist service personnel in IF problem isolation, however, an
instrument will function if the suspect functional parameters are not
required. This routine adjusts amplitude parameters first, then
resolution bandwidths in this sequence: 300 kHz, 1 MHz, 2 MHz, 100
kHz, 30 kHz, 10 kHz, 3 kHz, 1 kHz, 300 Hz, 100 Hz, 30 Hz, 10 Hz, 3 Hz,
and 1 Hz. The routine restarts from the beginning if a fault is detected.
Resolution bandwidth filters adjusted after the routine begins and
before the fault is detected should be acceptable; resolution bandwidth
filters adjusted later in the sequence are suspect. If the fault is detected
before the first resolution bandwidth (300 kHz) is adjusted, all IF
performance is suspect.
AMPL<300 Unable to adjust less than 300 Hz RES BW
ERR 400
amplitude.
AMPL 300 Unable to adjust amplitude of 300 Hz RES BW.
ERR 401
AMPL 1K Unable to adjust amplitude of 1 kHz RES BW.
ERR 402
AMPL 3K Unable to adjust amplitude of 3 kHz RES BW.
ERR 403
AMPL 10K Unable to adjust amplitude of 10 kHz RES BW.
ERR 404
RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 405
RBW 10K Unable to adjust 10 kHz RES BW pole 2.
ERR 406
RBW 10K Unable to adjust 10 kHz RES BW pole 3.
ERR 407
RBW 10K Unable to adjust 10 kHz RES BW pole 4.
ERR 408
RBW 10K Unable to adjust 10 kHz RES BW pole 1.
ERR 409
RBW 10K Unable to adjust 10 kHz RES BW pole 2.
ERR 410
Chapter 9
If You Have a Problem
Error Messages