Figure 2-54 Adjacent Channel Power Measurement Results - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Access the adjacent channel power (ACP) softkey functions and set up
the measurement parameters.
1. Press
power menu of softkeys.
2. Press
that will be used to make the measurement. Press
if it is not already selected. The analog method is the most common
measurement method and is described in more detail following this
example. Press
3. Press
channel bandwidth softkeys. Press
20.0 kHz. Press
PREV MENU
4. Then press
menu of functions.
Now that the setup is complete, you can activate the measurement.
5. Press
channel power measurement.
measurement parameters, optimizing the frequency span, resolution
bandwidth, video bandwidth, and detector of the spectrum analyzer.
Numerical results for the ACP leakage ratio in the upper and lower
adjacent channels are listed in the active function area, along with
the channel spacing and channel bandwidth. The total transmitted
power measured, and the maximum ACP ratio (of the upper and
lower adjacent channels) are listed in the title area in the
upper-right corner of the display.
Figure 2-54
Adjacent Channel Power Measurement Results
Chapter 2
, then
MEAS/USER
ACP MENU
and
ACP SETUP
METHODS
PREV MENU
SPACING/BANDWDTH
CHANNEL BANDWDTH
to return to the setup menu.
to return to the adjacent channel power
PREV MENU
ACP AUTO MEASURE
Example 9: Adjacent Channel Power Measurement
to access the adjacent channel
to select the measurement method
.
to access the channel spacing and
CHANNEL SPACING
and enter 14.0 kHz. Press
to activate the automatic adjacent
ACP AUTO MEASURE
Making Measurements
ANALOG METHOD
and enter
reconfigures the
111