3. Actuator (v3.1.1)
□
Lower Spreader bar to lowest possible position. (Lower limit)
□
From Kiosk or Tool Laptop on SafeGait network, open VnC and connect to the actuator (IP
address 10.0.0.2). Alternatively, open SafeGait cover to access HMI.
□
Log into Service Interface with service code 8
□
Navigate to the System Status Page (default)
□
Write down or take a picture of the ON time and RUN time for each axis:
Horizontal ON time
□
Navigate to Factory Page Load Cell Calibration
□
Write down or take a picture of the load cell bias and load cell coefficient for both axes.
Horizontal BIAS
□
Exit the service interface by clicking/tapping on the HOME PAGE button.
□
Connect an Ethernet cable from the HMI to the to the tool laptop
□
Navigate to where you copied the Actuator Installation Package (V3.0.0)
Option 1:
□
Open folder Configuration Values
□
Run the script: Config_Param.bat (read calibration and run time values)
□
On the command prompt type R to read values
□
Let script run
Horizontal RUN time
Horizontal COEF.
53
Vertical ON time
Vertical BIAS
SG360-ISM-004 | REV F
Vertical RUN time
Vertical COEF.