Alcatel ASM 180 TD User Manual page 85

Helium leak detector
Table of Contents

Advertisement

Ticket printing examples
(AUTO mode)
14/27
Alphanumeric Control and Display
Panel (ACDP) operation
FILE # : 1
CUSTOMER reference
S=level (mbar.l/s)
T=timer (sec)
S rejected : 2.8E-09
S alarm :9.9E-10
T roughing
: 9
T test
: 6
PRINTER ON
number: TEST #
26JUNE 96
CUSTOMER reference
TEST Nr : 5
AUTO MODE /FL
PASS signal = 9.0E-10
26JUNE 96
16.6:37
CUSTOMER reference
TEST Nr : 6
AUTO MODE /FL
PASS signal = 1.5E-09
Warning level exceeded
26JUNE 96
16.10:08
Modifiable and
validated SERIAL Nr
at the end of each test
CUSTOMER reference
TEST Nr : 7
AUTO MODE/
FAIL T. Roughing
26JUNE 96
16.12:05
Alcatel Vacuum Technology France - ASM 180 TD/TD+ - ASM 181 TD+ User's Manual
FILE # : 2
S=level (mbar.l/s)
T=timer (sec)
S rejected : 1.0E-08
S alarm :8.0E-09
T roughing
: 10
T test
: 7
PRINTER ON
number: SERIAL #
26JUNE 96
FILE # : 2
SERIAL Nr : 3826
AUTO MODE /FL
PASS signal = 7.2E-09
26JUNE 96
15.18:20
TEST Nr automatically
incremented
FILE # : 2
SERIAL Nr : 4217
AUTO MODE
FAIL signal = 3.2E-06
26JUNE 96
15.20:31
C 50
Parameters ticket
Selected part
Customer reference
Test ticket to be printed
Test reference
Test ticket
Accepted part
Accepted part with
warning level
Rejected part
Rejected part on
account of roughing
timer