Precautions

Access level 1

Cover securing
screws
Front cover
Rear cover
Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User's Manual
Access to internal detector components
ASM 182 TD+
Switch the detector off before opening it.
Disconnect the line power plug, the RS 232 connector, the
printer and the input/output interface if they are connected to
an external element (to avoid obtaining "external" voltages on
the various ASM 182 boards).
The front and rear covers are secured with 4 recessed screws.
Unfasten the 4 securing screws and their washers using a
5
5 mm Allen wrench. Use the 5 mm Allen wrench you find in
the "Maintenance kit".
Front cover:
After unfastening the 4 screws, raise the cover (1) slightly and
pivot it forwards until it comes to a stop (2).
Using this system, the cover can be kept 3/4 open.
Rear cover:
When opening the rear cover, take care with the ground
connection linking the frame to the cover.
If necessary, remove the front cover and/or the rear cover.
For the rear cover, you need before frese the rocker bar from
the frame.
1
2
Note: A hinge system is provided on the ASM182 TD+ for
the rear cover.

Maintenance kit

2
Rocker bar
Cover securing
screws
E 20
F 10
2
1/4

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ASM 182 TD+ and is the answer not in the manual?

Questions and answers

Summary of Contents for Alcatel ASM 182 TD+

  • Page 1 Cover securing screws Rocker bar Cover securing screws Front cover Rear cover Note: A hinge system is provided on the ASM182 TD+ for the rear cover. Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual...
  • Page 2 Release the assembly to the rear of the detector, taking care with the flat cable connectors attached to the I/O board. CHc (allen head) screw Front cover Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual...
  • Page 3 Remove the inlet air tube. Remove the internal calibrated leak. Disconnect the primary pump from the vacuum block (4 screws). Disconnect the MDP 5011 from the vacuum block (5 screws). Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual...
  • Page 4 Access to internal detector components ASM 182 TD+ Remove the 3 attachments screws of the vacuum block. Remove the 2 attachment screws of the arch obturator. Alcatel Vacuum Technology France - ASM 182 TD+ - ASM 192 T2D+ User’s Manual...

Table of Contents