Distributing the grease
in the bearings
2/2
Greasing the hybrid
turbomolecular pump PTM 5154
Set the SW1 switch on the CD2 board to the break-in (R)
position and run the detector for 10 minutes with it in this
position. Then reset SW1 to its normal (N) position.
SW1 switch :
R : Running-in
N : Normal
Alcatel Vacuum Technology France - ASM 180 TD/TD+ - ASM 181 TD+ User's Manual
E 20