Connecting the leak detector to
the installation via the hardware interface
Connect the remote control unit
1
The I/O interface connector should never be
Connect the
2
connected or disconnected with the unit on.
I/O interface
(see B 20)
If the detector is not
controlled by
the I/O interface:
the jumper plug
must be connected.
Connect the RS232
3
If the detector is to be connected to a micro-computer,
(see B 30)
connect the cabled RS232.
Connect an external
(ACDP option required see A 60 )
4
printer
(see B 40)
Connect the printer output using an RS232 cable.
Connect the LDS probe
5
(quick connector)
Connect to
6
When a neutral gas is used, the filter is unscrewed and
atmospheric pressure
replaced by the connection to the selected gas supply source.
1/4" BSP connector - Pressure : 1.0
Connect accessories
Additional plugs : 3 power plugs allow to connect accessories
7
such as recorders, gauges,...
(Maximum current: 4A; Specific 4A Fuses provided).
Connect the unit to
8
Check that the voltage marked on the unit
the main power
identification plate corresponds to that of the
electrical source .
Power consumption ...........................1.6 kVA
Fuse
Voltage 200-220-240 V .......3.15AT
Voltage 100-115 V..............6.30AT
Alcatel Vacuum Technology France - ASM 180 TD/TD+ - ASM 181 TD+ User's Manual
B 60
ASM 181 TD+ console version
It the detector is controlled
by the I/O interface, install
the interface cable to the
Sub D 25 pin connector on
the detector.