Finishing a measurement

You can finish a measurement in three ways:
1. You stop when the upper tensio shaft has reached the cavitation phase
(see illustration 1). Then you do not use the air entry point.
2. You are willing to use the air entry point. Then there are two possibilities:
a) The air entry point of the first tensio shaft has been reached and the second
tensio shaft is still in the regular measurement range (phase 1) or in boiling
delay (phase 2). In this case HYPROP can calculate the medial value
of the Tensio top and Tensio bottom curve (see illustration 2).
b) When the air entry point of the first tensio shaft has been reached and
the second tensio shaft is still in the cavitation phase (phase3), the medial
value of the two cannot yet be calculated. Then please wait until the air
entry point of the second tensio shaft has been reached (see illustration 3).
tension
[hPa or kPa]
0
Illustration 1
Tensio Top
Tensio Bottom
64 | Finishing a measurement
Stop
time