Adjusting

For precise measurements the balance needs to be adjusted to the local
conditions:
when it has been set up initially
after a change of position
after a temperature change.
We recommend adjusting the balance also every 4 weeks when the balance
is in measuring operation. For its adjustment the balance is equipped with an
adjusting weight.
Remove magneto cable and HYPROP
sensor unit.
Connect balance to the power supply
and turn it on.
Push and hold button until S.A.
CAL appears.
Push both buttons at the
same time and release them.
The readout flashes.
The zero point is saved.
55 | Preparing the measurement
Func
S.A. CAL
VaIT
CAL. 0
CAL. on
CAL
CAL