Offset recalibration

In the course of time the offsets of the electronic HYPROP sensors may slightly
drift. When you click the function "Offset Recalibration" the device checks
automatically if the offset values are within the narrow range set by UMS. Are
they are out of range the device recommends to run an offset recalibration.
Note
The offset recalibration can be done before every measurement as a routine to prepare
the sensor unit.
Fill and degas sensor unit (please find
description in the previous chapter).
Carefully remove sensor unit adapter, so
that all the water from inside is collected
on the up side of the sensor unit.
Connect sensor unit to the USB adapter.
Click the "Offset Recalibration" icon.
41 | Preparing the measurement
sensor
unit