Service Record - Welch Allyn Connex Service Manual

Integrated wall system
Hide thumbs Also See for Connex:
Table of Contents

Advertisement

Service manual

Service record

Print the Service record pages to enter and save the results of your tests.
Date:
Device name:
Technician:
Leak test
Leak test
Overpressure test
Overpressure test
NIBP accuracy check
Target pressure ± 5 mmHg
0 mmHg
50 mmHg
150 mmHg
250 mmHg
Leak test:
Over Pressure Test:
Pressure meter
Time:
Serial number:
Service tool version:
Specification
Max: 5
Specification
Adult mode:
280 . . . . 329
Neonate mode:
130 . . . .164
Service tool
mmHg
mmHg
mmHg
mmHg
mmHg
mmHg
mmHg
mmHg
Appendices 211
Actual reading
Pass
Actual reading
Specification
Pass
± 1 mmHg
± 3 mmHg
± 3 mmHg
± 3 mmHg
Fail
Fail

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents