Yearly Calibration
The INOblender calibration should be checked every 12 months using the NO monitor of the primary NO
delivery system. The test points should include all the control settings at oxygen gas flows of 5 and 14 L/min.
If the INOblender is outside of the accuracy specification, it should not be used.
Contact Customer Support at the number on the back cover of this manual for repair.
Replacing the CGA 626 tip on the INOMAX regulator
Note: • Depressurize the INOMAX regulator by using the purge port on the back of the INOmax DS
to removing from the cylinder valve.
• Check the INOblender and connection hoses for signs of wear and damage.
• For the CGA-type regulator connector, ensure the white plastic tip is not chipped or cracked. Remove
and replace as necessary. (see Figure 5-1).
b
a
Figure 5-1
Part No. 20732 Rev-02
2014-02
1. Disconnect the regulator from the INOMAX
gas cylinder.
2. Remove the old CGA 626 tip (for ISO see
Figure 5-2) by pulling on the tip (a) and turning
it counterclockwise (b) (see Figure 5-1).
3. Ensure the threads are clean on the
regulator tip (if required, use a lint free cloth).
4. Install the new tip:
Flex the four prongs by squeezing two prongs
at a time using only your fingers. This will help
start the new tip into the threads. Turn the tip
clockwise when threading the tip. When the
tip is fully inserted, it should turn freely.