Section 8: Specifications; Performance; Alarms - Mallinckrodt Nellcor N-20PA Service Manual

Portable pulse oximeter
Table of Contents

Advertisement

SECTION 8: SPECIFICATIONS

8.1

PERFORMANCE

8.1.1
Range
8.1.2
Accuracy
8.2

ALARMS

8.2.1
Alarm Limit Range
1
The reference for accuracy testing is an N-200 oximeter and D-25 sensors that have been validated in
human blood studies against an Instrumentation Laboratories CO-Oximeter. Adult accuracy specification
is based on testing with D-25 sensors; neonatal accuracy specification is based on testing with N-25
sensors. For specifications with other NELLCOR sensors, see the sensor Directions for Use.
2
This variation equals one standard deviation (SD). Plus or minus one SD encompasses 68% of the
population.
8.1 Performance
8.2 Alarms
8.3 Electrical
8.4 Environmental
8.5 Physical
Saturation
0–100%
Pulse Rate
20–250 beats per minute (bpm)
SpO
1
2
Adults
70–100% ± 2 digits
0–69% unspecified
Neonates
70–95% ± 2 digits
0–69% unspecified
Pulse rate
20–250 bpm
± 3 bpm
Saturation
20–100%
Pulse Rate
30–250 bpm
2
2
2
8-1