Prior to operation, complete the Pre-Use Checkout in Section 3.
The purge procedure contained in the Pre-use Checkout helps ensure any NO
regulator, INOblender and hoses before the manual resuscitator bag is connected to the patient.
Note: Depressurize the INOMAX regulator by using the purge port on the back of the INOmax DS
to removing from the cylinder valve.
2
3
1
1
2
The INOblender is now ready for patient use.
13
is purged from the pressure
2
1. Open the INOMAX cylinder valve.
2. Check for supply pressure of > 14 bar
(200 psig).
Adjust Settings
1. Turn the INOblender setting dial to the
desired concentration (5 to 80 ppm for an
800 ppm cylinder and 2.5 to 40 ppm for a
400 ppm cylinder).
2. Turn the O
flowmeter to the desired flow rate
2
(5 to 14 L/min).
3. Squeeze the manual resuscitator 3-4 times
to purge the NO
from the system.
2
prior
IR
Part No. 20732 Rev-02