Setup

2
3
To oxygen source
1
The INOblender is now set up. Proceed to Section 3/Pre-use Checkout.
1
2
7
Check the INOMAX gas cylinder for the correct
product identity labels, cylinder concentration
(800 or 400 ppm) and expiration date. Ensure the
INOMAX gas cylinder has > 14 bar (200 psig).
Note: • For the CGA-type INOMAX regulator
connector, ensure the white plastic tip
is not chipped or cracked. Remove and
replace as necessary. (see Replacing the
tip on the INOMAX regulator, Page 16).
• For the ISO-type regulator connector,
check that the O-ring is present and is
not damaged (see Replacing the tip on
the INOMAX regulator, Page 17).
1. Connect a high-pressure regulator to an
INOMAX cylinder and hand tighten the fitting
to the INOMAX cylinder.
2. Connect the INOMAX regulator hose to the
INOblender inlet hose using the quick
connect fittings.
3. Connect the oxygen supply to the O
fitting. (see part list for correct oxygen hose)
Note: 3.5 bar (50 psig) nominal
1. Make sure the NO dose setting dial is
turned to zero.
2. Attach a manual resuscitation bag to the
INOblender outlet fitting.
inlet
2
Part No. 20732 Rev-02
2014-02