4.2 Start-up Screens

At start-up, the type of screen displayed indicates the status of the system.
There are four common start-up screen types as follows:
System Startup
LOADING SYSTEM DATA
System Startup
NO SENSOR DETECTED
Device Resetting
System Startup
DUAL SENSOR MEMORY
Rectify & Reset
OI/FET100–EN Rev. A
System Start-up
At system start-up, a progress bar is displayed for the duration
of the start-up period.
After this period, one of the four following screens is displayed
according to the current status of the system.
No Sensor Connected
If no sensor is detected during start-up, an auto-recovery
routine is run to look for the sensor. If no sensor is detected,
this routine continues until it is stopped manually.
If 'Offline' is selected during auto-recovery (by pressing the
Offline
key) the transmitter ceases to operate as a flowmeter and
the following conditions apply at the transmitter:
Plant and transmitter data can be configured.
Sensor data cannot be configured.
Note. If this screen is displayed on an integral transmitter,
check that DIP switch SW3 is in the 'ON' position (refer to Fig.
3.13, page 18).
Dual Sensor Memory
Integral and retrofit systems have the sensor memory
mounted on the transmitter backplane board.
If two sensor memory types (integral and remote) are detected
at start-up, the warning 'DUAL SENSOR MEMORY' is
OK
displayed.
To correct this condition, set DIP switch SW3 to the 'OFF'
position (refer to Fig. 3.13, page 18).
4 Start-up and Operation
21