Establishing Communication; Testing The Transmitter - Siemens 7MF4033 series User Manual

Sitrans p ds iii series pressure, differential pressure, flanged level, and absolute pressure
Hide thumbs Also See for 7MF4033 series:
Table of Contents

Advertisement

UMSITRPDS3-1

3.2 ESTABLISHING COMMUNICATION

1. Connect the transmitter, power supply, and HART Communicator in a loop.
2. Apply power to the transmitter.
3. Press the HART Communicator's I/O key. The first display
is the Online menu (at right). Appropriate values for pressure
and device type should be displayed.
4. If the Online menu does not appear, or if a "Device not
found" message displays, check connections and power
source and repeat the above steps.

3.3 TESTING THE TRANSMITTER

Although a transmitter continuously performs an on-line self-
test, a more extensive self-test can be performed when
communication with the HART Communicator has been
established.
1. In the Online menu, choose option 3 Device setup, then press 2 Diag/Service, and finally press 2
Simulation/Test.
2. In the Simulation/Test menu, choose option 2 Test and then
press1 Self test or press the RIGHT ARROW/SELECT key
to start the test.
3. The transmitter performs the self-test.
If testing is successful, the message "Self test OK" appears and the Communicator returns to the
Test menu.
If testing fails, the message "Transmitter FAILED the
transmitter selftest" displays, and the transmitter goes to
the prescribed failsafe condition. Refer to Section 7
Calibration and Maintenance to troubleshoot the
transmitter.
4. Press the HOME softkey to return to the Online menu.
November 2005
Pre-Installation Test
SITRANS P
Online
1->Press
2 Type
3 Device Setup
HELP |SAVE
Note: If using a Model 375 Field
Communicator, see the inside of
the front cover of this manual.
3-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents