Printing Hard Copy Of The Monitor Screen; What You Need; Printer Setting; Signal Cable Connection - Kanomax Anemomaster 6162 Operation Manual

Anemometer
Hide thumbs Also See for Anemomaster 6162:
Table of Contents

Advertisement

2. Basic Operation

2.6 Printing Hard Copy of the Monitor Screen

2.6.1 What you need

★ Printer (sold separately) ····· Recommended model is DPU-201GS (Seiko Instruments Inc.)
★ Cable to connect the instrument and a printer (sold separately)

2.6.2 Printer Setting

Switch No.
SW 1
SW 2
SW 3
SW 4~6
Baud Rate
4800

2.6.3 Signal Cable Connection

Anemomaster
Signal Pin No.
3 (orange)
1 (brown)
4 (yellow)

2.6.4 Operation Procedure

1. Connect the instrument (RS-232C output terminal) and a printer.
2. Turn on both of the instrument and the printer.
3. Confirm that the Anemomaster's display is an initial screen.
4. Press HOLD key to hold the display screen.
5. When the display is on hold, press PRINT key.
To Halt Printing Temporarily and To Cancel Printing
Temporary Halt: To stop printing temporarily, press PRINT key while printing.
To recommence printing, press PRINT key again.
Cancel Printing: To stop printing temporarily, press PRINT key while printing.
When the printing is halted, press MENU key.
As the menu screen will be displayed, select 1. MONITOR.
Function
Anemomaster
Word Length
Parity Yes/No
Parity Setting
Baud Rate
SW 4
OFF
Signal Name
Signal Pin No.
TXD
GND
CTS
Printer
8 bit
ON
None
ON
None
ON
4800
Table Below
SW5
SW6
ON
OFF
Printer
Signal Name
3
DATA
5
GND
8
BUSY
'08.07.01
hold
10:26:40
5.38 m/s
25.7 °C
Press PRINT key when this
screen (on hold) is displayed.
13