Errors; User Interface; Title Screen; Main Screen - Kanomax 3950 User Manual

0.1um particle counter
Table of Contents

Advertisement

Errors

The following errors can occur by the self-diagnostic function. Even if an error occurs, the
measurement and other processes will be continued.
Laser power failure
Flow rate
Maximum
measurable
concentration
is exceeded

Title Screen

Turn on the power to the device. After four (4) seconds, the Main screen will appear
automatically. Or tap the Title screen to go to the Main screen.

Main Screen

There are three main screens as required. You can go to the sub-menu screens from the
main screens.
Table Form
Σ
Cumulative
The laser diode may be defective or reached the end of its life.
The pump current is increasing or decreasing. If the inlet
Protective Cap is attached, remove the Cap. If the error
persists, the pump may be defective or reached the end of its
life.
The concentration exceeds the maximum measurable range.
Perform the measurement at a cleaner location or attach the
Zero Filter to the device to measure.

User Interface

7