Kanomax Anemomaster 6162 Operation Manual page 41

Anemometer
Hide thumbs Also See for Anemomaster 6162:
Table of Contents

Advertisement

5. Data Output
(1)
# # ; # # # ; # # ; # # # ; # # # ; # # # # # CRLF
(1) Measurement Mode (VI: Air Velocity & Air Temperature)
Calculation Mode
(2)
(3) S-TIME (S) (Sampling Time)
(4) DATA (N) (Number of Data to be collected)
(5) POINTS (No. of samplings (partitions))
(6) Interval Measurement: INT(measurement interval)
Flow Measurement: AREA (m2) (Duct's cross section area)
(2)
[Output
(1)
Example]
VT ; FLW ; 60 ; 100 ; 10 ; 1. 456CRLF
# # # # # ; # # # # # ; # # # # # CRLF :19 Characters
air velocity
[Output
23.6;
Example]
<Command Explanation>
Command
To check the number of
P
memory pages
To check the number of
*
N
memory data
To check measurement
*
M
To output measurement
*
T
* When page number is entered incorrectly, the error message E will be returned.
As for E1 and E3, refer to 5.2.3 To Transfer Raw Data (measurement data per second).
Output Format of Measurement Condition ----- M Command
(2)
(3)
(4)
(AVE: Average Measurement INT: Interval Measurement FLW: Flow Measurement)
(4)
(3)
(5)
Output Format of Measurement Data ----- T Command
air temp
126;
Function
After the command is received, AP will be returned.
Then, the number of pages stored will be output.
After the command is received, AN will be returned.
Then, the number of data in the specified page (*) will be output.
After the command is received, AM will be returned.
condition
Then, the measurement condition in the specified page will be output.
After the command is received, AT will be returned.
data
Then, the measurement data in the specified page will be ouput.
(6)
(5)
(6)
dummy
: (Unit: m/s, ˚C, **)
:air velocity, air temp, dummy
*** CRLF
: 25 Characters
* AVG Mode:
dummy output
Explanation
35

Advertisement

Table of Contents
loading

Table of Contents