I/O Card Configuration - Magtrol DSP7000 User Manual

High speed programmable dynamometer controller dsp7000 series
Table of Contents

Advertisement

Magtrol Model DSP7000 Dynamometer Controller
pin
signal
name
14
DAC 1
common
15
DAC 2
common
16
AIN1-
17
AIN2-
18
5 Volts
Common
19
5 Volts
Common
20
5 Volts
Common
21
5 Volts
Common
22
DOUT2
23
Alarm
Relay1
Common
24
5 Volts
Common
25
Alarm
Relay2
Common
8.1.3
i/o C
arD
1. Press ShIFT.
2. Press SETUP button.
input/
level
output
N/A
Bidirectional Analog
N/A
Bidirectional Analog
Input
Bidirectional Analog
Input
Bidirectional Analog user
N/A
0 VDC
N/A
0 VDC
N/A
0 VDC
N/A
0 VDC
Output
OPEN or 8
Volts
Bidirectional
N/A
N/A
0 VDC
Bidirectional
N/A
C
onfiguraTion
Chapter 8 – Optional Equipment
intended use
secondary
common for
torque output
common for
speed output
User DAC 1
tachometer
common
input common
User DAC 2
input common
common
Supply
external sensor
common
External alarm
DIN3 common Switch closer to
common
DIN1 common
DIN2 common
Open collector,
less than
20 mA, can
be used to
switch on a
optocoupler or
small relay.
Relay output,
General
to shut down
purpose user
user equipment
relay
if DSP7000
is in an alarm
condition
Spare
Relay output,
General
to turn motor
purpose user
power on or off
relay
(Controlled by
MTEST)
89
notes
use
---
Signal return for
PIN 1
---
Signal return for
PIN 2
Signal return for
PIN 3
Signal return for
PIN 4
---
---
PIN 6
---
Switch closer to
PIN 7
---
Switch closer to
PIN 8
---
If inductive loads
are used, clamp
with a diode.
Recommended
used as a pilot
relay. 24 VDC, 1
amp max.
---
---
Recommended
used as a pilot
relay. LOW
VOLTAGE, LOW
CURRENT

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents