Anton Sprint eVo User Manual page 11

Multi-function flue gas analyser
Table of Contents

Advertisement

Sprint eVo 1 /eVo 2 /eVo 3
Note: for personal safety a carbon monoxide (CO) alarm will activate at 300ppm. This
will deactivate when CO levels drop below 150ppm. This is to protect the user from
potential hazardous exhaust gas levels.
To end test press the ESC key.
2. Let-by and Tightness test (eVo
Let-by test
Before carrying out the test, the pressure must be zeroed with the tube con-
nected to the instrument but not the pressure source. Note: in these tests 'Diff
press' is the difference between start and finish pressures.
1. Select Pressure Menu from the Test Menu. From the Press Menu screen
options select Let-by / Tightness.
2. Connect tube to positive pressure inlet but not pressure source.
3. Zero pressure, use the soft key marked
4. Connect tube to pressure source to begin pressure test.
5. Press the soft key
the screen.
6. To stop test press the soft key
7. Press the soft key
Let-by
14:44:27
08-Feb-2006
Start P1
mBar
Finish P2
mBar
Diff press
mBar
Duration
00:00
Screen 1
You may fail the test by pressing the key
(you can print the results).
Stabilisation Test
1. Press the soft key
the screen.
2. To stop test press the soft key
3. Press the soft key
to start test. Sprint displays the duration time on
u
.
to pass test and proceed to Stabilisation test.
ü
Let-by
14:44:27
08-Feb-2006
0.0
Start P1
0.0
mBar
Finish P2
0.0
0.0
mBar
Diff press
0.0
0.0
mBar
Duration
00:25
P=0
Screen 2
to start test. Sprint displays the duration time on
u
.
to proceed to Tightness test.
ü
Performing tests
2
3
and eVo
only)
.
P=0
Let-by
14:44:27
08-Feb-2006
Start P1
0.0
mBar
Finish P2
0.0
mBar
Diff press
0.0
mBar
Duration
00:54
Passed?
Screen 3
to indicate test has failed
û
9