Appendix I: Carbon Monoxide Room Safety Tests

Appendix I: Carbon Monoxide
Room Safety Tests
Sprint instruments are designed to assist heating engineers to work to the BS
7967 specification for carbon monoxide room safety testing. You should refer
to BS7967, which defines the requirements, details the methods as well as the
pass and fail criteria for various types of appliances.
CO Room Safety Test in the Sprint instruments is designed to measure the build
up of carbon monoxide levels in a room where a gas appliance is in use and
record those values each minute for the duration of the test. In addition at the
end of the test it assists the engineer (in an advisory capacity only) to determine
whether the test has passed or failed or if the results are invalid. In certain
circumstances, where the results are borderline or open to interpretation, the
instrument will ask the operator to decide if the test has passed or failed, and
will record the operators decision.
Please note: ultimately it is the responsibility of the operator to ensure that the test
is correctly performed to the BS 7967 specification. If the data does not support the
result or the operator suspects it is not reliable due to local conditions (such as carbon
monoxide level changes due to cigarette smoke or vehicle traffic) or incorrect, then
either the test should be repeated or the operator should seek expert advice.
CO Room Test Pass and Fail Test Specifications
Type C:
Room sealed
appliance
Max Allowed CO:
10 ppm
Max Peak Duration
60 secs
exceeding Max
Allowed CO:
CO Alarm Level:
30 ppm
Min Test Duration:
15 mins
Max Test Duration:
30 mins
Result Codes
The pass or fail result is displayed when the test completes and is printed on
the report as well as recorded in the log. If the test fails a code number is also
displayed, printed on the report and recorded in the log. This fail code identi-
fies the way in which the test failed and can help identify the cause. Also when
28
Type B:
Type A:
Type A:
Boiler
Cooker
Water heater
(open flue)
(flueless)
(flueless)
10 ppm
30 ppm
10 ppm
60 secs
20 mins
30 secs
30 ppm
90 ppm
30 ppm
15 mins
20 mins
5 mins
30 mins
30 mins
10 mins
Sprint eVo 1 /eVo 2 /eVo 3
Type A:
Space heater
(flueless)
10 ppm
60 secs
30 ppm
30 mins
30 mins