Anton Sprint eVo User Manual page 15

Multi-function flue gas analyser
Table of Contents

Advertisement

Sprint eVo 1 /eVo 2 /eVo 3
Performing tests
the CO Room Safety screens. The test can be completed by pressing the
soft
key to finish. Use the soft keys to log or print the results.
To stop test at any time, press the ESC key. Use the soft keys
or
to select
ü
û
'Quit test?'.
When printing Room CO safety test results Sprint will also provide data show-
ing for how long the recommended level of CO was exceeded. This is expressed
as hh:mm > allowed.
2
3
6. Gas escape test (eVo
and eVo
only)
Note: Battery power will be used at a higher rate under this test.
Select Gas Escape Detection from the Test Menu. Connect the Gas escape probe
(GEP) to the USB connetor, the light will illuminate on the probe. Sprint will
display a warning if the probe is not connected.
Sprint will display 'sensor stabilising' for approximately 30 seconds.
When the sensor is stable the unit will ask 'in clean air?' before zeroing.
Press
to zero the unit. Sprint will display a bar graph on the screen as gas
ü
levels are monitored.
Place the probe in the area of inspection for several seconds before moving it
to other locations.
Sprint will emit continuous clicks like a Geiger counter. If higher gas levels are
detected the bar graph will increase in readings and the sounder will increase
in pitch.
Press ESC to quit the test.
As with many types of instrumentation the gas escape probe may be damaged
by impact. If the probe is dropped or suffers significant impact in another way
check its operation by first plugging it into the Sprint and confirming that it is
recognised by the analyser.
If this is successful select the gas escape mode and briefly apply a sample of gas,
for example from an unlit cooker ring. This will allow the user to verify whether
the sensor is working correctly or not. If the gas is not detected contact Telegan
or your local distributor to arrange for the probe to be serviced.
It is recommended good practice that the above check is carried out periodically
even if the user is not aware of the probe being exposed to impact.
If the Sprint detects a fault with the GEP it will display a warning advising that
the leak sensor is faulty and the test will be aborted.
13