Error Codes

Under certain conditions, the gauge will display error messages. These
message are cleared by pressing any key. The cause of the error should
be corrected before proceeding.
Message
#1 - Probe
#2 - Probe
#3 - Probe
Probe Unsupported
Probe Changed
Unstable Reading
Value Too Large
Calibration Error
Bluetooth PC Reply
Error - EEPROM Write Gauge EEPROM communication error.
69
For more information
/
1300 287 935
R
Cause / Action
Probe to gauge communication failure.
Integral Gauges: Contact Elcometer.
Separate Gauges: Remove probe and refit. If error
persists, contact Elcometer.
Corrupt data output from probe.
Integral Gauges: Contact Elcometer.
Separate Gauges: Replace probe. If error persists,
contact Elcometer.
Internal Error - contact Elcometer.
No probe connected. Connect probe.
Probe is Dual FNF but gauge is ferrous or non-ferrous
only or Mk 3 probe is being used with a Mk 4 gauge.
Use correct probe.
Probe has been changed during a calibration
procedures. Calibration adjustment required.
External electromagnetic interference or gauge is faulty.
Isolate from electromagnetic interference. If error
persists, contact Elcometer.
Calculated value is too large to be displayed. Contact
Elcometer.
Incorrect calibration calculation. Re-calibrate. If error
persists, contact Elcometer.
Bluetooth link working at limit of its range. PC is busy
running other applications. Interference. PC
application doesn't send replies. Move gauge closer to
PC. Shutdown other applications. Remove source of
interference. If error persists, contact Elcometer.
If problem persists, EEPROM is faulty.