Getting Acquainted with Your Instrument
Timing Control
You can set parameters for sampling rate or clock using the Timing page of Settings window.
1. Sets Sampling Rate.
2. You can set Repetition Rate when the
Run mode is not set to Sequence.
3. You can select Clock Source (Internal
or External).
If External is selected, the clock
signal from External Oscillator Input
is used.
If Internal is selected, the clock signal
is generated internally.
4. Divider Rate can be set when the Clock
Source is set to External.
5. You can select the Reference Source
(Internal or External).
Reference Source is selectable only
when the Clock Source is set to Internal.
6. You can select the External Reference
Type (Variable or Fixed).
This parameter is selectable only when
the Clock Source is set to Internal and
the Reference Source is set to External.
Quick Tips
The Sampling Rate can be set when the internal clock source is selected and one of the following conditions is met:
Internal is selected as the reference source.
External is selected as the reference source and Fixed is selected as the external reference type.
The arbitrary waveform generator accepts a 10 MHz, 20 MHz, or 100 MHz frequency signal as a Fixed external
reference source.
You can set the Multiplier Rate when the Clock Source is set to Internal, the Reference Source is set to External, and
the External Reference Type is set to Variable.