8. The copied waveform is pasted to
index 2.
9. Click the Switch to Table icon to change
the display to Tabular view.
10. Set the following sequence parameters:
Index 1: Wait Trigger On
The sequence waits for a trigger
before generating the first element.
Index 2: Repeat count Infinite
The sequence infinitely generates
the Sine-1 waveform until an event
occurs.
Index 2: Event Jump To Next
When the instrument receives an
event signal, the sequence jumps to
next element.
Index 3: Repeat count 1000
The sequence repeats the
*Triangle960 waveform 1000
times.
Index 4: Go To 1
The sequence jumps to index
number 1 after generating index
number 4 waveform.
11. Click the Run button to confirm that the
sequence works properly. The animation
window next to the Run button will be
activated when the instrument is in the
running state.