Ultrasound Troubleshooting - GE Corometrics 126 Service Manual

Corometrics 120 series
Table of Contents

Advertisement

Ultrasound Troubleshooting

Problem
Ultrasound not functioning properly.
Static noise on ultrasound.
Rate on FHR area of display and FHR
trend on strip chart paper do not correlate.
Revision B
Troubleshooting: Ultrasound Troubleshooting
Table 15-2. Ultrasound Troubleshooting
Probable Cause
Transducer not properly connected to
monitor.
Transducer placement.
Too little gel applied to transducer.
Defective transducer.
Active fetus or mother. Fetal arrhythmia
or hiccups. Extreme maternal obesity.
No signal.
Service required.
Active fetus.
Environmental noise.
Maternal movement.
Defective transducer.
Monitor is set for 30 BPM/cm vertical scale
and 20 BPM/cm vertical scale strip chart
recorder paper is being used (or vice
versa).
120 Series Maternal/Fetal Monitor
2015590-001
Possible Solution
Ensure that transducer is securely
attached to monitor.
Wait before moving transducer; FHR
often returns. Reposition transducer.
Apply more gel.
Replace transducer.
Use alternate technique.
Auscultate FHR.
Call Biomedical Engineering
Department.
Reposition transducer.
Keep sheets and gown off transducer.
Do not hold transducer with hand.
Use alternate monitoring mode.
Replace transducer.
Call Biomedical Engineering Department.