Table 4.1 Configuration instruction
Setting
Button
Dock due warning
Dock due (Sync) interval
Maintenance interval
Description
Button effect
This display screen features the status, alarm, configuration, warning, and "dOC", icons, with
the selection number in the main area. The technician can disable, or enable the warning, and
choose the warning type.
When enabled, the unit will notify the user that docking to a station is due, based on the
selected option. The instrument will continue to operate.
Note: When options 1, 2, or 3 are selected, the expected battery life will be reduced.
Values:
0 = disabled
1 = enabled for audible chirp
2 = enabled for blue LED flash
3 = enabled for combination audible chirp and blue LED flash
4 = enabled for display only
Increments the value.
One press saves the displayed value and activates the next configuration-mode screen.
This display screen features the status, configuration, calendar, and "SYnC" icons, with the
current value in the main area. The technician can set the interval at which the dock due
warning is to be activated.
Note: When there is a high or low alarm, the unit will display the dock due icons automatically.
Value range: 1 to 365 days
Value increment: 1 day
The first press activates the value. Continued presses increment the value; hold to speed the
increment pace.
One press saves the displayed value and activates the next configuration-mode screen.
This display screen features the status, clock, configuration and warning icons, with "min" in
the lower left area, and the time interval setting in the main area. The technician can choose
the signal frequency for all enabled calibration, bump test, and dock due notifications.
For example, if the technician sets the interval for five minutes, every five minutes the
indicator selected from the enabled maintenance due warnings (calibration, bump test, and
dock due) will be activated.
Value range: 1 to 60 minutes
Value increment: 1 minute
The first press activates the value. Continued presses increment the value; hold to speed the
increment pace.
One press saves the displayed value and activates the next configuration-mode screen.