Table 4.1 Configuration instruction
Setting
Button
STEL operation-mode
STEL alarm setpoint
Calibration gas
Time
Description
Button effect
This display screen features the status, lock, configuration, and STEL icons. The technician
can enable or disable the option for operation-mode access.
When enabled, the instrument operator is permitted to view and clear the unit's STEL reading
while the unit is in operation mode.
Values:
0 = disabled
1 = enabled
Increments the value.
One press saves the displayed value and activates the next configuration-mode screen.
This display screen features the status, alarm, sensor type, configuration, and STEL icons,
with the current setpoint. The alarm setpoint can be edited.
Value increment: sensor measurement resolution
See Table 1.5 for more information about the installed sensor type.
The first press activates the value. Continued presses increment the value; hold to speed the
increment pace.
One press saves the displayed value; a second press activates the next configuration-mode
screen.
This display screen features the status, sensor-type, configuration, and calibration icons, with
the current calibration gas setting in the main area.
This setting reflects the concentration of calibration gas that the instrument expects to read
when calibrated; it should be edited to match the cylinder's gas concentration.
Value range: within the sensor measurement range
Value increment: sensor measurement resolution
See Table 1.5 for the measurement range and resolution for each sensor type.
Increments the value; hold to speed the increment pace.
One press saves the displayed value; a second press activates the next configuration-mode
screen.
This display screen features the status, clock, and configuration icons, with the current time
setting.
The instrument's clock uses a 24-hour time format. Its settings are edited in this order using
these values:
Hours: 00 to 24
Minutes: 00 to 59
Value increment: 1
The first press activates the first value to be edited. Continued presses increment the value;
hold to speed the increment pace.
One press saves the displayed value and activates the next value to be edited.
Continue to use the buttons,
After all values are edited and saved, one press activates the next configuration-mode screen.
and
, to edit and save the values, respectively.