Table 4.1 Configuration instruction
Setting
Button
Operation-mode calibration
Calibration due warning
Calibration interval
Calibration date
Description
Button effect
This display screen features the status, lock, configuration, and calibration icons. The
technician can enable or disable this operation-mode option.
When enabled, the instrument operator is given access to calibrate the unit from operation
mode.
Values:
0 = disabled
1 = enabled
Increments the value.
One press saves the displayed value and activates the next configuration-mode screen.
This display screen features the status, alarm, configuration, calibration, and warning icons.
The technician can disable or enable the warning, and choose the warning type.
When enabled, the unit will notify the user that calibration is due, based on the selected
option. The instrument will continue to operate.
Note: When options 1, 2, or 3 are selected, the expected battery life will be reduced.
Values:
0 = disabled
1 = enabled for audible chirp
2 = enabled for blue LED flash
3 = enabled for combination audible chirp and blue LED flash
4 = enabled for display only
Increments the value.
One press saves the displayed value and activates the next configuration-mode screen.
This display screen features the status, configuration, calibration, and calendar icons, with the
current setting in the main area. The technician can set the interval at which the calibration
due warning is to be activated.
Value range: 1 to 365 days
Value increment: 1 day
The first press activates the value. Continued presses increment the value; hold to speed the
increment pace.
One press saves the displayed value; a second press activates the next configuration-mode
screen.
This display screen features the status, up and down arrow, configuration, calibration,
warning, and calendar icons. The technician can choose whether the operation-mode
calibration date screen will display the due date for the unit's next calibration or the date of the
unit's last calibration.
The up arrow (▲) will be featured on-screen when the unit is set to display the next the
calibration due. The down arrow (▼) will be featured when the unit is set to display the last
calibration date.
Values:
0 = displays date of last calibration
1 = displays next calibration due date
Increments the value.
One press saves the displayed value and activates the next configuration-mode screen.