Conversion Standards – UCI
The default measuring method is according to ASTM A1038 and DIN 50159.
For conversions, the user has the choice to select either ISO or ASTM.
Impact direction (Leeb only)
With exception to DL and U devices all Leeb impact devices have auto-
matic direction compensation. You may override this and set the impact
direction manually. For more information on impact direction please refer
to chapter "3.1.1 Leeb Testing Procedure (except Leeb U)". Impact direc-
tion is not relevant to the Portable Rockwell and the UCI devices.
Trigger Load (UCI only)
For the Equotip UCI probe the load where the measurement will be trig-
gered can be chosen in the range from HV1 to HV5 (10 – 50 N). The trig-
ger load cannot be changed once the measurment series was started.
Units (Portable Rockwell only)
For Portable Rockwell probe, choose to display the indentation depth in
metric or imperial units.
4.1.4
Sample IDs
After Measurement
Use this setting to define if the current sample ID's should be kept for the
next measurement series or deleted.
Edit Entries
The entries of the different sample ID fields can be deleted or edited here.
For easy increasing or decreasing, use the up- and down- arrows. For add-
ing or removing entry fields, please see chapter "8.1.2 General Features".
4.1.5
Workflow
Activate User Guidance
Select to display on screen instructions and messages when taking a
measurement.
26
Use Advanced Algorithm (Portable Rockwell only)
Advanced Algorithm provides faster measurements. This is especially
useful when testing softer material.
Auto Close Series
Automatically end a series after a set number of measurements. The user
can set the series from 1 to 1000 measurements.
Measurement Comment Handling
Use this setting to allow or disallow the user to enter a comment at the
completion of a measurement series. When set to "free" this enables the
user inputing a comment.
Measurement Series Filename
Enter the file name for the measurement series will be stored. This pos-
sibility is disabled, if filename management is activated.
Save to Folder
Set the folder location where the measurement series file to be stored.
This option is disabled, if folder management is activated.
Store Signal Data (Leeb only)
Select to store the raw waveform for Leeb measurements. For Portable
Rockwell the signal form will be stored for each measurement automatically,
for UCI this option is not available..
NOTE! Storing signal data will cause measurement files to
take up more memory.
Enable Warnings
Choose to enable warning display signals and sounds to indicate false
measurements.