CE APPROVAL INFORMATION
1. Electromagnetic Compatibility (EMC)
This device comforms with requirements of EMC Directive 89/336/EEC,
amended by 93/68/EEC. This instrument complies with the following
EMC Immunity Standards as tested per EN 50082-2, 1995 (Industrial
environment)
Phenomena
Test Specification
Electrostatic
+/- 4 kV contact discharge
Discharge
+/- 8 kV air discharge
Radio Frequency
27 - 1000 MHz
electromagnetic
10 V/m
field.
80% AM (1 KHz)
Radio Frequency
900 MHz
electromagnetic field.
10 V/m
Pulse modulated.
50% Duty cycle @ 200 Hz
Fast Transients
+/- 2 kV (ac mains)
+/- 1 kV (dc, signal I/O)
5/50 ns Tr/Th, 5 KHz rep. freq.
Radio Frequency
0.15 - 80 MHz
conducted
10 V/m
80% AM (1 KHz)
94
This instrument complies with the following EMC Emission Standards
as tested per EN 50081-1, 1992 (Residential, Commercial and Light
Industrial)
Phenomena
Basic Standard
Radiated
IEC 1000-4-2
Emission
Performance
Criteria B
Conducted
IEC 1000-4-3
Emission
Performance
Criteria A
IEC 1000-4-3
2. Safety
Performance
This device conforms with Low Voltage Directive 73/23/EEC, amended by
Criteria A
93/68/EEC. The following LVD requirements have been met to comply with
EN 61010-1, 1993
IEC 1000-4-4
(Electrical equipment for measurement, control and laboratory use)
Performance
Criteria B
1. Pollution Degree 2
2. Installation Category II
IEC 1000-4-6
3. Double Insulation
Performance
4. Class I Equipment (90-240 Vac Powered Units)
Criteria A
Frequency
Limits
Range
30-230 MHz
30 dB_V/m at 10 m
230-1000 MHz
37 dB_V/m at 10 m
quasi peak
0.15-0.5 MHz
66-56 dB_V quasi peak
0.5-5 MHz
56 dB_V quasi peak
5-30 MHz
60 dB_V quasi peak
Basic
Standard
CISPR 22
Class B
CISPR 22
Class B