Kodak DryView 8150 Safety Manual page 8

Laser imager
Hide thumbs Also See for DryView 8150:
Table of Contents

Advertisement

September 27, 2004
CAUTION
This equipment should not be located within the patient
environment. Therefore, do NOT locate this equipment
closer than 1.83 meters (6 feet) from a patient bed or chair.
1.83 m
CAUTION
This equipment should not be used in close contact with MRI
devices, due to possible very high magnetic fields near an
MRI unit. The magnetic field in the area where this
equipment is installed must be less than 50 Gauss.
CAUTION
Do not substitute or modify any part of this equipment
without prior written approval of Eastman Kodak Company.
CAUTION
(USA Only)
Federal law prohibits dispensing without a prescription.
CAUTION
FCC (USA Only)
This equipment has been tested and found to comply with
the limits for a class A digital device, pursuant to part 15 of
the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the
7F3779
Safety and Related Information
:
:
:
:
:
1-3