Kodak DryView 8150 Safety Manual page 20

Laser imager
Hide thumbs Also See for DryView 8150:
Table of Contents

Advertisement

Safety and Related Information
Recommended separation distance between portable and mobile RF communications
equipment and the Kodak DryView Model 8150 Laser Imager
The Kodak DryView 8150 Laser Imager is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Kodak DryView 8150 Laser Imager can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication
equipment (transmitters) and the Kodak DryView 8150 Laser Imager as recommended below, according to the
maximum output of the communications equipment.
Rated maximum output
Separation distance (d) according to frequency of transmitter
power of transmitter (P)
in meters (m)
in Watts (W)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d
=
1.17 P
d
=
2.33 P
d
=
1.17 P
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.