Write reader outputs Command (F2
RF Carrier on/off Command (F4
Writes the state of the reader outputs. The flags are ignored for this command.
The data section contains 1 byte with bits defined as follows:
Bit 0
1 = Output 1 switched on (output is pulled to ground)
Bit 1
1 = Output 2 switched on (output is pulled to ground)
Bit 2
Reserved
Bit 3
Reserved
Bit 4
1 = Bit 0 enabled (output 1 is controlled)
Bit 5
1 = Bit 1 enabled (output 2 is controlled)
Bit 6
Reserved
Bit 7
Reserved
Example
Switch output 2 on without affecting output 1.
Request packet
01 0A 00 00 00 00 F2 22 DB 24
The response packet is similar to the request packet with the data section containing
'00
' for a successful write operation.
hex
Example
Response packet
01 0A 00 00 00 00 F2 00 F9 06
Write successful.
)
hex
Switches the RF carrier on or off. The data section contains one byte FF
carrier on or 00
to turn the carrier off.
hex
Example
Switch the carrier on.
Request packet
01 0A 00 00 00 00 F4 FF 00 FF
The response packet is similar to the request packet with the data section containing
'00
' for a successful operation.
hex
Example
Response packet
01 0A 00 00 00 00 F4 00 FF 00
Command successful.