Kodak DryView 8150 Safety Manual page 18

Laser imager
Hide thumbs Also See for DryView 8150:
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration for Electromagnetic Immunity
The Kodak DryView 8150 Laser Imager is intended for use in the electromagnetic environment specified below. The
customer or user of the Kodak DryView 8150 Laser Imager should assure that it is used in such an environment.
Immunity
IEC 60601 Test
Test
Level
Conducted RF
3 Vrms
150 kHz to 80 MHz
IEC 61000-4-6
Radiated RF
3 v/m
80 MHz to 2.5 GHz
IEC 61000-4-3
NOTE:
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
See Note 1 on next page.
b
See Note 2 on next page.
September 27, 2004
Compliance
Electromagnetic Environment - Guidance
Level
Portable and mobile RF communications equipment should be
used no closer to any part of the Kodak DryView 8150 Laser
Imager, including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended Separation Distance:
3 Vrms
d
=
1.17 P
3 v/m
d
=
1.17 P
d
=
2.33 P
d is the recommended separation distance in meters (m).
P is the maximum output rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
compliance level in each frequency range
Interference may occur in the vicinity of equipment marked
with the following symbol:
7F3779
Safety and Related Information
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a
, should be less than the
b
.
1-13

Advertisement

Table of Contents
loading