4 CALIBRATING THE GAUGE (CONTINUED)
4.3 CALIBRATING WHEN USING DIFFERENT CALIBRATION TILES
If using a different tile to the one identified by the Elcometer 407L, the calibration values of the tile
must be manually entered, before performing the calibration procedure.
1
Select Menu/Setup/Calibration Menu
2
Press the
or
3
Follow the on screen instructions to set the tile value
4
Repeat for "Tile Value 60" and "Tile Value 85", where appropriate.

5 TAKING A READING

1
Place the gauge on the substrate and press the
2
The reading is taken and the measurement data updated
The gauge can be set to take automatic readings at 2, 5 or 10 second
intervals. This feature can be enabled via Menu/Setup/Measurement
Menu/Auto Measure.
Pressing the
navigation button will access the gloss statistics screen,
displaying statistics for the current batch. Readings can also be taken in
this screen and the statistics will be updated accordingly.
Pressing the
navigation button in the gloss statistics screen will
access the gloss graph; this gives a graphical representation of the gloss
measurements.
Note: Graphs (dependent upon the gloss angle available) will only be displayed when two
or more measurements have been taken.
6
navigation button to highlight "Tile Value 20" followed by
navigation button
R
to select