3.5 SETTING THE PASS/FAIL PARAMETERS (CONTINUED)
When enabled, any measurement which falls between the min and max values will be considered a
pass and the header text will be coloured green.
Readings outside the min and max values will be deemed a fail and the header text will be coloured
red.

4 CALIBRATING THE GAUGE

The gauge has an automatic calibration option which detects when the Elcometer 407L calibration tile
is in position. This is switched ON by default but can be disabled via Menu/Setup/Calibration Menu.
The gauge is supplied with filter paper protecting the calibration tile. This must be removed before
calibrating.
4.1 CALIBRATING WHEN AUTOMATIC CALIBRATION IS ENABLED
1
Place the gauge on the calibration tile and press the
message will be displayed:
"
Tile Detected -
Note:
The calibration values of the tile supplied with the Elcometer 407L are pre-programmed into the gauge. If
using a different tile, follow the procedure outlined in Section 4.3.
2
P
ress the
or
3
Press the
or
the calibration
4.2 CALIBRATING WHEN AUTOMATIC CALIBRATION IS DISABLED
1
Place the gauge on the calibration tile and select Menu/Calibrate Instrument
2
Press the
or
the calibration
Note: The calibration values of the tile supplied with the Elcometer 407L are pre-programmed into the gauge. If
using a different tile, follow the procedure outlined in Section 4.3.
b
Triple Angle gauge only.
www.elcometer.com
Assigned Values; 20° - xx.x; 60° - xx.x; 85° - xx.x
navigation button to highlight "CALIBRATE" followed by
navigation button to highlight "CONFIRM" followed by
navigation button to highlight "CONFIRM" followed by
Dual Angle gauges will display 20°- xx.x; 60°- xx.x.
navigation button. The following
b
.
"
to select and save
to select and save
to select
5